INTRODUCTION
The instrument is a precision AC energy meter testing instrument, mainly used to test three phase energy meter error on site and measure various of AC parameters.
TECHNICAL SPECIFICATIONS
Electrical parameters
Accuracy class 0.05%, 0.1%
Resolution 6 TFT (640-480)
Power supply 220V±10%, 50/60Hz
Li-polymer battery (size (mm): 110x51x16, nominal output voltage: 7.2V, capacity: 5000mAh)
Power line supply (U1, UN), 85V-265V 50/60Hz
Communication port RS232
Test voltage
Range 5-480V
Error ±0.05% (30V-480V)
±0.1% (5V-30V)
Harmonic 2nd-63st
Test current
Range (direct connection) 5mA-20A
Error (direct connection) ±0.05% (100mA-20A)
±0.1% (10mA-100mA)
Range (clamp CT) 10mA-3000A
Error (camp CT) ±0.2% (100mA-100A)
±0.5% (100A-3000A)
Harmonic 2nd-63st
Power measure error
Active power (direct connection) ±0.05% (0.1A-20A)
±0.1% (0.01A-0.1A)
Reactive power (direct connection) ±0.1% (0.1A-20A)
Energy measure error
Active energy (direct connection) ±0.05% (0.1A-20A)
±0.1% (0.01A-0.1A)
Reactive energy (direct connection) ±0.1% (0.1A-20A)
Phase angle
Range 0-360
Resolution 0.01
Error ±0.05
Frequency
Range 45-65Hz
Resolution 0.001Hz
Error 0.002Hz
Pulse input
Input channel 2
Input level 5-24V
Input frequency Max. 2MHz
Pulse output
Energy constant 180000imp/kWh, 1800imp/kWh, 180imp/kWh
Pulse ratio 1:1
Output level 5V
Function
Vector diagram Yes
Waveform Yes
Energy accumulation Yes
Communication with PC Yes
Mechanical parameters
Instrument dimensions (WHD) (mm) 220-13861
Instrument Weight (kg) 1.7
Carry case dimensions (WHD) (mm) 450-320185
Carry case (kg) 8.5
Environmental conditions
Ambient temperature -10C to 55C
Relative humidity 15%-85%