Olympus converges its latest advancements in high-performance digital circuitry and eddy current flaw detection into one compact and durable portable unit the new NORTE 600. With its crisp and vivid 5.7 inch VGA display and true full-screen mode, the NORTEC 600 is capable of producing highly visible and contrasting eddy current signals in any lighting condition.
The redesigned interface of the NORTEC 600 borrows the intuitive, knob-operated navigation of its NORTEC predecessors and combines it with the simple menu structure and highly efficient direct-access keys of other popular Olympus instruments. Available in four versatile models, the NORTEC 600 offers a wide range of innovative functionalities, including an Application Selection menu, an all-in-one display, real-time readings, and signal calibration in Freeze mode, ensuring that inspections are quick and easy for any level of operator.
Reliable and Robust
Based on an already field-proven instrument case, the NORTEC 600 is built for endurance in harsh field conditions. Its casingâ??s durability and resistance to extreme weather makes the NORTEC 600 an instrument you can trust for any eddy current inspection job.
Key Features
Designed to meet the requirements of IP66.
EN-15548 compliant.
Long battery life (up to 10 hours).
Bright, 5.7 inch (14,5 cm) VGA display.
Full-screen option in any display mode.
Improved filters for rotary scanner mode.
Intuitive interface with Application Selection presets.
All-settings configuration page.
10 Hz to 12 MHz frequency capacity.
Automatic internal balancing (BNC connector).
Up to two real-time readings.
True automatic mixing.
Storage capacity of up to 500 files (program and data).
On-board file preview.
Includes:
- Olympus NORTEC 600D Dual Frequency
- Calibration certificate
- Charger/adaptor
- AC power cord
- Getting Started leaflet
- NORTEC 600D Users Manual and PC interface program on CD-ROM
- Universal PowerLink eddy current probe cable
- MicroSD memory card, 2 GB (1 internal, 1 external)
- USB 2.0 (mini-AB) to PC cable
- 67 watt-hour battery for 600 Series products, 10.8 V, 6.2 Ah, 67 Wh
- AA battery holder for emergency use
- Factory-installed hand strap on left side of NORTEC 600D instrument
- Instrument transport case
- Olympus NEC-8196L Weld Inspection Kit
Specifications
PRF 10 Hz to 2000 Hz in 10 Hz increments
Pulse Width Adjustable from 25 ns to 5,000 ns (0.1 MHz) with PerfectSquare technology
Gain 0 to 110 dB
Receiver Bandwidth 0.2 MHz to 26.5 MHz at â??3 dB
System Linearity Horizontal: �± 0.5% FSW
Resolution 0.25% FSH, amplifier accuracy �± 1dB
Reject 0 to 80% FSH with Visual Warning
Amplitude Measurement 0 to 110% full screen height with 0.25% resolution
Automated Calibration Velocity, Zero Offset
Straight Beam (First Backwall or Echo-to-Echo)
Angle Beam (Soundpath or Depth)
Test Modes Pulse Echo, Dual, or Through Transmission
Range 3.36 mm to 13,388 mm (0.132 in. to 527.10 in.) at 5,900 m/s (0.2320 in./�µs)
Velocity 635 m/s to 15240 m/s (0.0250 in./�µs to 0.6000 in./�µs)
Zero Offset 0 to 750 �µs
Refracted Angle 0�° to 90�° in 0.1�° increments
Measurement Gates 2 fully independent gates for amplitude and TOF measurements
Measurements
Gate (1, 2) Thickness, Soundpath, Projection, Depth, Amplitude, Time-Of-Flight, Min./Max. Depth, Min./Max. Amplitude
Echo-to-Echo Standard Gate 2-Gate 1, Optional IF Gate Tracking
Other measurements Overshoot (dB) value for DGS/AVG, ERS (equivalent reflector size) for DGS/AVG, AWS D1.1/D1.5 A, B, C and D values, Reject Value, Echo to RefdB values
DAC/TCG Standard
DAC points Up to 50 points, 110 dB dynamic range
Special DAC modes Custom DAC (up to 6 curves), 20-80% View
Curved surface correction Standard OD or Bar correction for Angle Beam measurements