Test Method:Dispersion light nephelometric analysis
Test Principle:The highest point on first order differentiation,Percentage analysis
Zero point correction:auto track the zero point,eliminate the interference of sample and reagent
Wavelength:470nm
Light Source:Cold Lamp with high brightness and long lifetime
Detector:unique Dispersion light detector with high sensitivity and repeatability,immunized the effect on result of jaundice¡¢hemolysis¡¢chyle and turbidity on plasma
Test Channel:each channel is independent,one or more,same or different items can be tested Simultaneously
Incubator:Provide reagent and sample with stable incubator,
Incubator Temperature:37¡À0.2oC
Measure trigger:Auto trigger the measurement through configured pipette
Item modify:easy to add or edit items,easy to set the items parameters
Calibration:single-point calibration;multipoint calibration;auto store and match with calibration curve
Quality control:auto draw the quality figure
Reagent:open reagent system
Reagent volume:40
Sample volume: 40
Memory:Save more than 10000 samples results,and auto recovering the results when power is off during working
Repeatability:CV¡Ü5%
Calculative item:S¡¢%¡¢PTR¡¢INR,Fib can be get through PT,This leads to lower the cost of reagent
Printer:Inner thermal printer,can be connected to external printer
Information edit:easy to input and edit the detail information of patients
Interface:RS-232 serial port,can be connected to external printer
Working Environment:Temperature:10 oC~30 oC Humidity: ¡Ü70%RH
Voltage:85-264V,Auto adapt according to the input voltage
Power: ¡Ü80W
Quality System:Recognized by CE,TUV and CMD
Test Items
PT
AT-¢ó
APTT
Protein-S
TT
Protein-C
Fib
LA
RepT
Hep
VT.
260mm*257mm*133mm
3.6kg.