INTENDED USE Feline Immunodeficiency Virus Ab Test is a sandwich lateral flow immunochromatographic assay for the qualitative detection of Feline Immunodeficiency Virus antibody (FIV Ab) in cat's blood. PRINCIPLE OF THE ASSAY Feline Immunodeficiency Virus Ab Test is based on sandwich lateral flow immunochromatographic assay. The test device has a testing window. The testing window has an invisible T (test) zone and C (control) zone. When sample is applied into the sample hole on the device, the liquid will laterally flow on the surface of the test strip. If there is enough FIV antibody in the sample, a visible T band will appear. The C band should always appear after a sample is applied, indicating a valid result. By this means, the device can accurately indicate the presence of FIV antibody in the sample. INTERPRETATION OF RESULTS Positive: The presence of both C band and T band, no matter T band is clear or vague. Negative: Only clear C band appears. Invalid: No colored band appears in C zone, no matter whether T band appears. STORAGE The kit can be stored at room temperature (2-30C). The test kit is stable through the expiration date (18 months) marked on the foil pouch. DO NOT FREEZE. Do not store the test kit in direct sunlight.