With different external probes, the gauge can be applied to measuring thickness of non-magnetic coating on magnetic metal substrate, as well as non-conductive coating on non-magnetic metal substrate. Five types of probes (F400, F1, F10, N1, CN02) are available for different application. F10 probe measures up to 10mm. Two measuring modes: single or continuous, changeable. Two calibration methods can be applied to the gauge; And the system error of the probe can be corrected with the basic probe calibration method. Measured values and user information are shown on a large, easy-to-read LCD. A display back light ensures easy reading of screen data in poorly-lit conditions. Measuring status indicator showing the measuring status. The gauges¡¯ user-friendly measuring system permits automatic storage of up to 1000 readings in one memory matrix for later statistical evaluation. Measuring Principles The gauge adopts two thickness-measuring methods: magnetic induction method and eddy current method. insulating coatings on all non-ferrous metals and on austenitic stainless steels, e.g. paint, anodizing coatings, ceramics, etc. applied on aluminum, copper, zinc die-casting, brass, etc. Working Conditions Working temperature£º -10¡æ¡«£«50¡æ; Storage temperature£º-30¡æ¡«£«60¡æ; Relative humidity: ¡Ü90£¥; The surrounding environment should avoid of vibration, strong magnetic field, corrosive medium and heavy dust. Configuration No. Item Quantity Remarks Standard Config. 1 Main unit 1 2 Probe 1 F1 or N1 3 Calibration Foils 5 PCS 4 Zero plate 1 Iron or Aluminum 5 Instrument case 1 6 Manual 1 7 Alkaline Battery 2 PCS AA size 8 Optional Config. 9 Other type of probes 10 DataPro software 1 On CD 11 USB cable 1