Ventilation Mode:
IPPV. A/C, PCV , SIMV, Spont/CPAP, PSV, Sigh, Manu
Ventilator Parameter
Tidal volume:0, 20- 2000 rnL
(Vt. Frequency (Freq):1 min - 100 min
Oxygen concentration: 21 % - 100 %
I:E:4 : 1 -- 1:8
PEEP:0 cmH2O 40 cmH2O
Pressure limit: 20 cmH2O -100 cmH20
Pressure triggering sensitivity(Ptr): -20 cmH2O - 0 cmH2O(Based on PEEP)
Flow triggering sensitivity (Ftr): 0.5 cmH2O - 30 cmHz0
SIGH: 0 (off) 1/100 - 5/100
Apnea ventilation: OFF, 5s - 60s
MONITORING PARAMETER
Frequency (Freq): 0/min - 100/min
Tidal volume (Vt): 0mL - 2500mL
MV: 0 lUmln - 99 Umin
Airway pressure: 0 cmH2O - 100 cmH2O
Dynamic lung compliance monitoring:1mL cmH2O - 1000 mL/cmH2O
Oxygen concentration: 15 % - 100 %
ALARM & PROTECTION
AC power failure alarm: Power failure or no connection
Internal backup backup battery low voltage alarm: <11.3 �± 0.3 V
No tidal volume:No tidal volume within 6s
High minute volume alarm:5L/min - 99L/min
Low minute volume alarm:1L/min - 30L/min
High airway pressure alarm: 20 cmH2O - 100 cmH2O
Low airway pressure alarm: 0 cmH2O - 20 cmH2O
High oxygen concentration alarm:19 % 100%
Low oxygen concentration alarm:18 % - 99 %
Continuous pressure alarm: (PEEP + 1.5 cmH2O) over 16s
Suffocation warning: 5-60s
Fan error: On Screen
Oxygen deficit: On Screen
Maximum limited pressure: < 12 kPa
WORKING CONDITION
Gas source: 02, Air
Pressure:280 kPa 600 kPa
Voltage:220 V �± 22V
Power frequency:50 Hz ± 1 Hz
Input power: 900VA (Air Compressor) - 250 VA (WIthout Air Compressor)
OSCILLOGRAM
P-T(Pressure-Time)
F-T(Flow-Time)
P-V Loop(Pressure-Volume Loop)