A rapid, one step test for the qualitative detection of Amphetamines in human urine. For healthcare professionals including professionals at point of care sites For professional in vitro diagnostic use only. A drug-free urine pool was spiked with Amphetamine at the following concentrations: 0 ng/mL, 500 ng/mL, 750 ng/mL, 1,000 ng/mL, 1,250 ng/mL and 1,500 ng/mL. The result demonstrates >99% accuracy at 50% above and 50% below the cut-off concentration. The data are summarized below: AMP Concentration (ng/mL)Percent of CutoffnVisual Result NegativePositive 00%30300 500-50%30300 750-25%30237 1,000Cutoff30921 1,250+25%30129 1,500+50%30030