Tidal Volume:20-2500ml
Display:10.4 TFT touch screen
Type:pneumatically driven tronic control
Gas supply:medical o2, medical air
Pneumatic system safe pressure:125cmH2O
Power supply:AC 110-240V, 50-60 Hz
Battery:Lithium battery, backup for more than 2 hours
Product Size:H400mm* W303mm * L250mm Weight: 16kg
HS code:9031809090
Real-time waveforms: Paw, Flow, Volume, EtCO2, SPO2
Optional: CO2, SPO2 (related parameters)
Loops: P-V, V-Flow, Flow- P
Trends: 1s, 5s,1min,10min, 30min, 60min
Tidal volume: 5 to 2000mL
PEEP/CPAP: 0 to 30 cmH2O
Oxygen: 21 to 100%
I:E ratio: 1:10 to 4:1
Inspiratory time (TI): 0.2 to 5 s
Flow trigger: 0.2 to 15 l/min
Pressure trigger: -0.5 to -20cmH2O
Pressure control: 5 to 90 cmH2O, added to PEEP/CPAP
Pressure support: 0 to 90 cmH2O, added to PEEP/CPAP
Pressure ramp: 0 to 2s
Expiratory trigger sensitivity (ETS): OFF, 5 to 90 % of inspiratory peak flow
Peak flow Spontaneous: > 210 L/min
Input voltage: 100 to 240V ~ �?�±10 %, 50 / 60 Hz AC or 24 VDC�?�±10 %
Power consumption: 50 W typical
Protection class: Class I
Backup battery: batteries(5200mAh each)/ hot swappable
Oxygen supply: 280 to 600 kPa (41 to 87 psi), 120 L / min
Low pressure: oxygen �¢?�¤15 L / min, max. 600 hPa for low pressure
Operator-adjustable: Low/high minute volume, low/high pressure, low/high tidal volume, low/high rate, apnea time, low/high oxygen
Optional: low/high CO2, low/high SPO2
Special alarms: Oxygen concentration, disconnection, loss of PEEP, exhalation obstruction, flow sensor, power supply batteries, gas supply
Loudness Adjustable: 1 to 9
Event log: Storage and display of up to 1000 events with date and time stamp
Weight: < 45.3 kg
Dimensions(WxHxD): Approx. 600x402x415mm
Power supply
Voltage: 100~240V
Frequency: 50HZ, 60HZ
Input Power: 80VA
Internal Power: DC12V�¯�¼? 4AH
Packing Size
Main engine: L 560 * W 560 * H 605 mm
G.W.:35kg
Air Compressor: L 683 * Wv687 * H 1140 mm
G.W.: 100kg