Suitable for Adult | Pediatric | Neonatal
The lung ventilator for Intense Care Units (ICU) LUFT3, counts with integrated 17" touch screen technology and offers a complete range of ventilation modes which allows the
monitoring of the patients condition.
Ventilation Modes:
**ADULT/PEDIATRIC**
VCV assisted/controlled
PCV assisted/controlled
PSV/CPAP
PRVC assisted/controlled
SIMV (VCV) + PSV
SIMV (PCV) + PSV
MMV + PSV
PSV + assured VT
Biphasic pressure (APRV + PSV)
NIV
HFNC
**NEONATAL**
VCV assisted/controlled
PCV assisted/controlled
PSV/CPAP
SIMV (PCV) + PSV
Continuous flow assisted/controlled
Nasal CPAP
HFNC
**SPECIAL FEATURES**
- 100 % oxygen up to 20 min. with automatic alarm silence;
- Adaptation of the patients interface or change of the circuit with recalibration without having to turn off the equipment and keeping track of the patients log;
- Automatic theoretical weight calculation and interface selection according to the patient;
- Altitude compesation;
- Configuration of the monitored variables;
- Flow curve 50 % descending;
- Intuitive Interface;
- Gas measurement with BTPS correction;
- Smart ventilator, records the user preference after 10 uses;
- LCD Screen tilt angle adjustment;
- Log for 1000 alarms and events with date and time.
**OPERATIONAL VISUALIZATION**
- Graphical pressure bar
- Indicator of spontaneous/controlled cycles
- Battery charge level
- Programation of the ventilatory variables
ADULT AND PEDIATRIC - 5 SIMULTANEOUS GRAPHICS
- Pressure/Time
- Flow/Time
- Volume/Time
- Volume/Pressure
- Flow/Volume
- Pressure/Flow
NEONATAL - 3 SIMULTANEOUS GRAPHICS
- Pressure/Time
- Flow/Time
- Volume/Time
**LUNG MECHANICS**
- AutoPEEP
- Slow vital capacity
- Dymanic compliance
- Static compliance
- PV curve with low flow
- Elastance
- Stress Index
- Tobin Index
- P0.1 - Airway obstruction pressure
- Expiratory resistance
- Inspiratory resistance
- WoB (Work of breathing)
PR4-g
Ventilation modes
Adult / pediatric
Assist-controlled VCV
Assist-controlled PCV
PSV / CPAP
SIMV (VCV) + PSV
Back-up ventilation (VCV)
Back-up ventilation (PCV)
NEONATAL
Continuous flow
Nasal CPAP
Programmable alarms
Maximum pressure
Minimum pressure
Maximum tidal volume
Minimum tidal volume
Apnea alarm time
Maximum frequency
Non-programmable alarms
I:E ratio inversion
PEEP
O2 supply fail
Interrupted cycle
Controls / Programmable parameters
FIO2: 50-100%
Rise time: 10-100 L/min
Inhalation time: 0.3 û 3 sec
Frequency: 1 û 150 cpm
Tidal volume: 10 û 1500 ml
Inspiratory pressure: 2 û 30 H2O cm
Support pressure: 2 û 30 H2O cm
PEEP: 0 û 20 H2O cm
Pressure trigger sensitivity: -0.5 to -10.0 H2O cm
Altitude compensation: 0 û 6000 msnm
Inspiratory pause
0.10 - 2 seg
Monitoring / output parameters
Peak airway pressure: 0 û 120 H20 cm
Mean airway pressure: 0 û 100 H2O cm
Inspiratory tidal volume: 50 û 1500 ml
Frequency: 1 û 150 cpm
I:E: 5:1 û 1:99
Graphic monitors
Pressure-time waveform
Waveform hold
Auto scale
Other features / functions
Historic¦s alarms
Lung mechanics: Auto PEEP