Function and usage
It can set with 5 different temperature spots to do heat sealing, select the optimal heat healing temperature, and can also set the same temperature to speed up the heat sealing.
Instrument structure:
The structure of the heat-sealing tester is composed of 4 parts: pneumatic drive system, heat-sealing equipment, temperature adjustment system and time adjustment system.
Pneumatic drive system:
It includes pressure gauge, guide stick, cylinder, pneumatic valve, pressure regulating handle and so on. You can pull the pressure-regulating handle outward, rotate to the right side to adjust the heat sealing pressure to a high value, and rotate to the left side to reduce it. After setting the pressure, please press the handle to the side of mainframe to lock the air pressure.
Heating-seal device
It is composed of the upper and lower heat-sealing sticks, of which the upper one is connected with a cylinder. There is an insulating plate in the upper part, which can prevent heat transmit to the cylinder. The heat-sealing stick is made of aluminum, and the center is heater. There is a silica gel plate in the upper part of the lower heat-sealing stick, which can mitigate pressure impact during heat sealing. Temperature measurement system was installed at the back of the heat-sealing stick.
Temperature adjustment system:
The upper temperature controller controls the temperature of upper heat-sealing stick, and the rear temperature controller controls the temperature of lower heat-sealing stick. The temperature can be set at will through adjusting the button on the dial plate.
Time adjustment system:
Time adjustment system: The heat-sealing time can be set at will through adjusting the button on dial plate. The setting scope of time is 0.01s to 999.9h. Press the foot switch under the manual state, and the heat-sealing stick will be pressed together. Under the automatic state, the press-fitting time of upper and lower heat sealing sticks are controlled by a timer automatically, it can automatically recover later. The unit of timer is 0.1 second in general. There is time adjusting button on the panel of the timer, and the heat-sealing time can be adjusted at will.
Specifications
Power : AC220V 50HZ
Compressed air: 8kg/cm2
Heat sealing plane: 300mm10mm slick (With silicon pad in lower seal stick)
Heat sealing temperature: room temperature 20~300C (Precision 1C)
Heat sealing time: 0.01s~99.99h
Heat-sealing pressure: 0~0.8Mp.
Dimension: 700(L)mm400(W)mm540(H)mm
Weight: Desktop 30Kg.
700mm(L)X400mm(B)X540mm(H)
Function and usage:
Gas chromatography carries out the quantitative analysis and peculiar smell testing for residual solvent on food packaging cases after printing;Test the pureness and quality of organic solvents.
Features
1. With higher ratio of function and reasonable price, economic and efficient
2. The design of the cellular integral structure is convenient for installation.
3. It can be connected to the chromatogram data work station so as to greatly improve the working efficiency.
4. With FID detector£ºHydrogen flame ionization detector for organic compounds with high sensitivity
5. With the flexible design of the packed column and capillary column injection ports.
6. Packed column and capillary column sample injectors can be selected as per actual situations. Wide column box (300, 300, 200mm)
7. In order to meet complicated analytic task, it makes use of various chromatographic columns so as to offer convenience and widen the range of the analytic samples. The advanced noise suppression system can fully eliminate the fan noise under the chromatogram working state, so as to provide you a comfortable and quiet working atmosphere.
8. The sampling device is completely vitrified with two modes of rapid vaporizing and column sample injection.
9. Bran-new pressure stabilization and various air circuit system
10. The air circuit has three different kinds of forms: Single, double and triple. It can match six-way valve, blow back, and precut equipment and so on according to the users¡¯ demands. It adopts new type pressure regulation system to ensure the re-productivity of equipment.
11. Reforming furnace can be equipped to implement microanalysis of CO and CO2.
12. Cracking furnace can be equipped to do double eight-step temperature rise with microcomputer control, man-machine conversation and simple application.
13. Advanced temperature control system
14. Adopt microcomputer control and man-machine conversation to indicate various needed temperatures of the temperature column box.
15. Column box temperature control scope: (room temperature +10C) - 400C, eight-step temperature rise.
The speed is 0 - 30C/min (increase 0.1C/min), and the temperature control precision is 0.1%.
The control point number is eight points of A type with control scope (room temperature +10C) - 400C.
16. Overheat protection function
It can freely set temperature within 400C with software program protection, and the hardwired connection adopts dual fail-safe design
72cm*70cm*70cm Gross weight:65KG.
Introduction
N530 for the detection of gas permeability of various films, infusion bags and other packaging materials for O2, CO2, N2 and other gases through the barrier properties. It is designed and manufactured according to the standards of ISO 2556-2001, ISO15105-1, GB / T 1038-2000, ASTM D1434-82 (2003), YBB00082003 .It has three independent testing chambers and independent temperature control system. The core is connected with computer through the USB. It can automatically judge and stop testing . The testing result is stable, curves of permeability rate ,temperature, pressure and different pressure will be shown at the same time. The test report will generate automatically while the data automatically stored. It also support TCP/IP protocol. There are three measuring range that can be enlarged.
Test principlesN 530 Gas Permeation Analyzer designed using the principle of differential pressure method, preparation samples will be placed between in the test chamber with high / low pressure , and clamping it ,then make the whole test system to have a vacuum processed, input test gas the high-pressure test chamber to ensure on both sides of the test samples have a constant pressure, the gas under the role of pressure pass through film from high-pressure chamber to the low pressure chamber to infiltrate, observe the changes pressure in low-pressure chamber to caculate barrier parameters of the sample.
Specifications
Measuring range: 0.02 ~ 50000 cm3/m2 day 0.1Mpa (Through the expansion of volume cap can test 600000 cm3/m2 day 0.1Mpa) Measurement Accuracy: 0.01cm3/m2 day 0.1MPa
Temperature range: 5 ~ 50 (standard temperature of 23 )
Temperature Accuracy: 0.01
Humidity Range: 0% RH (5 ~ 100% RH optional)
Wet control accuracy: 2% RH
Vacuum accuracy: 0.1Pa
Test pressure:-0.1Mpa ~ +0.1 Mpa
Test gas: O2, CO2, N2 and other gases
Air Pressure: 0.1Mpa ~ 0.8Mpa
Specimen size: 80mm, through an area of 50.24 cm2
Sample: 3(extension up to more pieces)
Carrier gas interface: 1 / 8 inch
Sample thickness: 2mm
Power: AC 220V /50HZ
Instrument weigh : 75 Kg.
Gloss weight: 90 Kg.
Dimension : 80 cm, 56 cm, 40 cm
Features
1. Computer automatically control three chambers of testing ,can test different samples, Configuration three independent sensors, each chamber can be set different temperatures. First creating positive and negative pressure dual-mode testing, leading international standard.
2.The sample can be tested by single chamber, or double chambers or three chambers .
3. Seperately show the curve of transmission, pressure, pressure value and temperature.
4. High or low levels of barrier test, repeat excellent performance.
5. The international advanced electromagnetic process step temperature control technology, ease lifting the temperature, variable temperature adjustable-rate, temperature control accuracy of 0.01 range.
6. High and low resistance ,two kinds of test mode7. Excellent test reproducibility.
8. An excellent environmental adaptability, no special requirements on the environment.
9. Real-time display of each point transmission, through the transmittance curve, can understand the testing process.
10. The curves of each chamber¡¯s temperature, humidity, weight, and transmission display independently, continous to display four- groups-curve, can monitor the interaction of parameters at the same time.
11. Software interface is simple and easy to operate, automatically test.
12. Overpressure protection13. Software interface is simple and easy to operate, automatically save the test report can be printed and other parameters and the transmittance curve of the report.
14. Optional pairs of air humidity controlling devices, more precise of
controlling wet;
15. Calibration method of standard film.
Technical Specifications:
Measuring range: 0.02~10000 gram/m 2 /24 hr
SENSOR : 0.0001g (best European brand, tested recommended by NASA)
Test Chambers: W 300 (one chamber and a sensor, W320 (Two chambers and two
sensor) W330 (three chambers and three sensors)
Testing precision: 0.001 gram/m2/24 hr
Testing temperature: 10~50
Temperature control precision: ±0.1
Humidity Range: 0%~30%RH (Best recommended as per International standards), (also available as per Chinese standard (5%- 90% RH)
Humidity precision: ±2%RH
Specimen Size: 80 mm*10cm
Silica gel required: 250g / Chamber
Testing area: 50.24cm2.
Wooden box.
Technical Specifications:
Measuring range: 0.02~10000 gram/m 2 /24 hr
SENSOR : 0.0001g (best European brand, tested recommended by NASA)
Test Chambers: W320 (Two chambers and two
sensor)
Testing precision: 0.001 gram/m2/24 hr
Testing temperature: 10~50 C
Temperature control precision: ±0.1
Humidity Range: 0%~30%RH (Best recommended as per International standards), (also available as per Chinese standard (5%- 90% RH)
Humidity precision: ±2%RH
Specimen Size: 80 mm*10cm
Silica gel required: 250g / Chamber
Testing area: 50.24cm2.
Wooden box.
Application
To test the water vapor transmission rate (WVTR) of packaging materials, such as plastic film, composite film, coextrusion film, aluminum-plated film, aluminum foil, infusion bag, sheets, solar battery panel, cellophane, ceramics and porcelain, and various containers such as bag, pouch, bottle, can, bowl, box, widely used in the industries of food, pharmaceuticals, personal care, household, electronics and so on.
Working Principle
Electrolytic detection sensor method. Fix the test sample in the middle of test chamber to separate the chamber into upper room and lower room. When humid gas flows in upper room and dry gas in lower room, the water molecules in upper room penetrate through the sample into the dry gas, and electrolytic sensor system detect and analyze the water content and calculate the water vapor transmission rate.