PLX101 Series
High Frequency Mobile X-ray Equipment
PLX101/101A/101C
Rocker safety lock design, make your operate much safer
High quality x-ray monoblock assures more details, less dose
Symmetrical adjustable collimator for accurate positioning and better exposure protection
Aviation aluminum connectors, connection reliable, more safety
Advanced dislocation braking system, flexible movement, flexible braking
Multifunctional cassette storage box design
Features
Our high frequency technology assures the best image in every operating condition
Along with reduced patient dose and extended x-ray tube life time
The generator provide the right amount of penetration, reducing exposure times and retakes.
The parameter can be adjusted by the kV and mAs. Multiple security protect function such as over-voltage, over-current.
50 preset exposure programs allow extensive possibilities to optimize the exposure and processing parameters according to your preferences.
Fault automatic protection setting and fault alarming display
Parameters are automatically storage when power cut off.
Adopt rotating beam limiting device to adjust the X-ray field and angle freely.
Adopt manual controller and remote controller to do exposure.
Optional: PLXF152 mobile table for high frequency mobile X-ray equipment.
PLX101D is equipped with built-in battery, can take 200 films when full charged.
Rocker can move up and down, automatically balance
High quality X-ray monoblock can rotate in ±90
Options
PLXF152 Mobile Bed For Mobile X-ray Machine
Size: 2000 X 700X 500mm
Range of grid box movement 1210mm
Size of grid box: 430X 430(17 X 17)
Bucky Stand
Height: 1770mm
Vertical movement range: 470-1470mm
Grid density: 103L/inch
Grid radio: 10:1
Focus distance: 150cm
Maximum grid box: 17 X 17
100 mA, 100 KVP Mobile X-ray Machine
100mA, 100 KVP, Mobile X-ray Machine for Radiography only
Electronic Timer
Electronic Overload Protection Device
Stationary Anode X-ray Tube : BEL or Imported
Collimator
Counter Balanced Mobile Stand
Suitable for Bed Side X-rays
We're a leading XRay machine and accessories manufacturer company with over 29 years of experience and global presence and distributor network in 7 different countries.
We have Line Frequency & High Frequency, APR X-Ray Machine, Dental X-Ray Machine, Mobile 100MA X-Ray Machine, 300MA X-Ray Machine, 500MA XRay Machine, Vertical Bucky, Multi-Position Bucky Table, 4 & 2 Way Floating Bucky Table Etc
Beside new machines, we offer complete modification & repairing of any brand XRay machines, Rotating Anode Tube Head, HT 100MA, 300MA, 500MA Etc
2.5kw high frequency portable X-ray machines PLX101
x ray tube Fixed Anode
Power Output of x ray machine 2.5kW
Frequency 25kHz
Tube voltage 40~100kV
Power Supply 220V�±10%, 50kHZ
Operation Method Wire/ Wireless control
price list of x-ray machine
price of 500ma x-ray machine
price of medical x ray machine
X-ray machine
Technical parameter
1. The numbers of x-ray tubes used in the product are also used as product number and x-ray generator (bulb tube head) number.
2. The current and voltage of x-ray tubes are fixed in three groups, they are:
X exposure voltageú¿kvpú¬ 60 65 70
X-ray currentú¿maú¬ 0.50 0.85 1.10
3. When adding up voltage, total current is less than 2a and total power is 500w
4. Drain current and earth resistance are in accordant with the stipulates in gb9706.1
5. The three groups of voltage (60kvíó65kvíó70kv) are adjustable.
6ú«wireless remote control in any angles within 10 meters part form the control box
7. The parameter of x-ray generator (bulb tube head)
7.1. The parameter of x-ray tubes
A. Focus index: 0.3í-0.3mm2
Bú«the voltage of the-highest-voltage generator: 90kv
C. The highest inverse peak voltage: 90kv
D. Filament characteristic: vf=5ví+0.5vú¼if=3a
E. Target surface obliquity: 12íp
F. The cooling manner of bulb tube head: oil immersion
8.7.2 the parameter of x-ray generator (bulb tube head):
A. Focus-skin distance: 200mm
B. The thickness of filter-ray film: more than 1.5mmal
C. The leakage rate of x-ray: no more than 0.002mgy/h (no. 29.204.3 in gb9706.12 prescript that the leakage rate of x-ray is no more than 0.25mgy/h)
D. The manner of limiting beam: using planigraphic device
Technical parameter
1. The numbers of x-ray tubes used in the product are also used as product numbers and x-ray generator (bulb tube head) numbers.
2. The current and voltage of x-ray tubes are fixed in three groups, they are:
X exposure voltage (kvp) 60 65 70
X-ray current (ma) 0.50 0.85 1.10
3. When adding up voltage, total current is less than 2a and total power is 500w
4. Drain current and earth resistance are in accordant with the stipulates in gb9706.1
5. The three groups of voltage (60kv, 65kv, 70kv) are adjustable.
6. Wireless remote control in any angles within 10 meters part form the control box
7. The parameter of x-ray generator (bulb tube head)
7.1. The parameter of x-ray tubes
A. Focus index: 0.3í-0.3mm2
B. The voltage of the-highest-voltage generator: 90kv
C. The highest inverse peak voltage: 90kv
D. Filament characteristic: vf=5ví+0.5v, if=3a
E. Target surface obliquity: 12íp
F. The cooling manner of bulb tube head: oil immersion
7.2 the parameter of x-ray generator (bulb tube head):
A. Focus-skin distance: 200mm
B. The thickness of filter-ray film: more than 1.5mmal
C. The leakage rate of x-ray: no more than 0.002mgy/h (no. 29.204.3 in gb9706.12 prescript that the leakage rate of x-ray is no more than 0.25mgy/h)
D. The manner of limiting beam: using planigraphic device
Tube stand (xts-01)
Type : floor-to-ceiling rail
Distance of longitudinal travel : 2550mm
Distance of vertical travel : 1600mm
Rotation of x-ray tube : ±180°
Lock system : electromagnetic lock by manual
Weight : 170kg
Ceiling (xce-01, option)
Type : ceiling rail
Distance of longitudinal travel : 2800mm
Distance of vertical travel : 1200mm
Rotation of x-ray tube : ±180°
Lock system : electromagnetic lock by manual
Weight : 130kg
Bucky stand (xbs-01, option)
Type : vertical travel
Distance of vertical travel : 120mm ~ 1780mm
X-ray grid : grid ratio of 8:1, 103 line/inch
Front plate to film distance : 50mm
Weight : 105kg
4way type bucky table (xbt4-01)
Type : 4way travel
Longitudinal distance of tabletop sliding : 420±5mm
Lateral distance of tabletop sliding : 120±5mm
Tabletop to floor distance : 665mm
X-ray grid : grid ratio of 8:1, 103line/inch
Dimensions : 2000l x 750w x 665h (mm)
Weight : 150kg
6way type bucky table (xbt6-01, option)
Type : 6way travel
Drive type : electric motor drive
X-ray grid : grid ratio of 8:1, 103line/inch
Longitudinal distance of tabletop sliding : 420±5mm
Lateral distance of tabletop sliding : 120±5mm
Vertical distance of tabletop : 300±5mm
Tabletop to floor distance : 689(880)mm
Dimensions : 2000l x 750w x 680(880)h (mm)
Weight : 170kg
High voltage cable (hvc-6/8)
Length : 6m/8m
Maximum voltage : dc 75kv
The number of core line : 3 lines