NC 654Performance Specifications Measuring Strokes : 600(mm) x 500(mm) x 400(mm) Overall Dimension : 1630(mm) x 1280(mm) x 2380 (mm) Max part weight : 500 kg Resolution : 0.1 (mm) Accuracy according GB/T16857.2-1997 equal ISO 10 360ª (Volumetric length measuring error) E =2.5 +3.5L/1000 [mm](Volumetric Probing Error) R = 2.4 mm Environment Specifications Room required specifications Average room temperature: 18 - 22 pC Maximum thermal gradients:1 C/h2 pC/day1 pC/m Air supply system (Min. air pressure) : 0.7 MPa Electrical requirements - (Voltage) : 220 Vac + 10% Frequency) : 50 Hz (Current) : 10 A Allowed Maximum Floor Vibration (Frequency Range): 0 ~ 19 Hz 19~75 Hz > 75 Hz